case history 5
Warming Product Heating Element
project

Medical Warming Product

objective

Our client wanted to design a very compact warming product for use in operating rooms in Europe. Marketing indicated that there would be very great acceptance of the device if it could be very small. The electronic 'noise' from in-rush current would be very objectionable so the heater needed to ramp up in stages. The heater and product had to be very safe.

challenges

The heater in this product would provide warm air into a plastic suit surrounding the patient during and after surgery. It was imperative that this compact product with plastic and heat operate safely in a sensitive environment like the Operating Room with oxygen nearby and the patient connected.

Since the product needed to be such a small size, there was very little room for the heater. A space of 1.8" (46mm) x 7.5" (190mm) x .8" (20mm) is all that was available for a heater in the most desired product configuration. To accommodate 1400 watts @ 240 volts and squeeze in safety devices would be difficult - especially with adequate electrical spacings for CE approval.

solutions &
their impact
  • Farnam was able to design a heater which would fit this space without compromising the needed rate of airflow. We designed the heater so that integral mica dielectric / thermal insulators would line the plastic duct and act as heat shields.
  • We implemented a design with three circuits which allows the heater to ramp up to 1400 watts in stages. The three circuits are powered up in a sequence of 200 then 400, and then 800 watts. The in-rush current is controlled by this gradual and progressive process.
  • A snap acting over-temperature thermostat and a back up thermal fuse are integrated in series with all three circuits. The heater is protected against excessive temperature build up.

 

 